Nihon Kohden Vismo: battery error with a new battery

What happened

After a new battery was installed, the monitor reported a battery error and the charge indicator flashed quickly.

What was checked

Cause

As reported in the case: after the battery swap, the battery reset (recognition) step had not been performed, so the monitor did not recognise the new battery and kept flagging a fault.

What was done

The fix was a procedure from the manufacturer's service documentation, done by authorised personnel: the battery reset was carried out and the monitor was restarted normally. Charging resumed and, as reported in the case, the charge indicator settled to a slow green light.

Safety

Before any battery work, disconnect the monitor from mains power and inspect the pack for swelling, heat or leakage — a damaged lithium battery is a fire risk — and verify battery backup and alarms before the monitor returns to clinical use.

Lesson: A battery error on a brand-new battery is often a recognition issue rather than a bad part, so run the manufacturer's battery reset procedure after replacement before ordering anything.