Nikkiso · DBB-007

Nikkiso DBB-007: cascading hydraulic valve alarms

What happened

As reported in the case, the machine raised valve errors in sequence — first on SV4, then SV8 — ending with a persistent SV5 alarm that would not clear.

What was checked

Cause

In this case the alarms came from a combination: old filters had been reused after drying out, apparently for cost saving, the acid and bicarbonate concentrate batch was stale, and on top of that some valves had genuinely failed — the contaminated hydraulic path triggered the sequence, and the valves needed replacing anyway.

What was done

The engineers replaced the failed valves, renewed all filters and both concentrate batches, and ran several rinse cycles; the machine passed its checks and the alarms did not return. Outcome: solved.

Safety

Treat the hydraulic path and used filters as biologically contaminated: wear gloves and eye protection, contain spills, and complete full disinfection with a leak check before the machine returns to clinical use.

Lesson: Cascading valve alarms on a dialysis machine often start with consumables and hidden leaks, so check filter condition, concentrate freshness and the connector lines before condemning the solenoids.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.