Nuve NC150: autoclave won't start, "system almost ready"

What happened

Reported in the case: the sterilizer refused to start cycles, displaying "system almost ready" together with intermittent "insufficient water" and "replace HEPA filter" warnings.

What was checked

Cause

The fill pump of the steam generator had failed. Without water the heater did not switch on, so pressure and temperature never rose and the controller kept the unit in the "not ready" state. The water and HEPA-filter messages were secondary effects, not the fault itself.

What was done

The fill pump of the steam generator was repaired/replaced. After that water filled normally, the heater started, pressure and temperature rose, and the sterilizer launched cycles. The case was reported as solved.

Safety

Before opening the water circuit, let the steam generator cool and release residual pressure, and after the repair run a full sterilisation cycle plus an electrical safety check before returning the unit to service.

Lesson: A "not ready" status is usually the end of a chain — verify the real prerequisites (water fill, heating, pressure rise) before chasing each individual warning message.