Olympus CF-Q150L: green tint and dark spots in the image after the scope was submerged
What happened
Reported in the case: after the video colonoscope had been submerged in liquid and then dried out, the image showed a green tint along the upper edge and dark spots across the lower part of the field.
What was checked
- The distal objective lens was cleaned and re-inspected in good light — it came back clean, so the artefacts were not on the outer glass.
- A leak test of the scope was recommended to confirm where the liquid got in and whether it is still getting in.
Cause
Probable cause discussed in the case: dried residue and stains under the objective lens. Liquid that entered the scope left deposits inside the optical assembly after drying, and that shows up as colour shift and dark patches in the image.
The cause is probable and was not fully confirmed.
What was done
No confirmed fix was recorded in the case, and the outcome is probable only. Contamination under the objective cannot be cleaned from outside; the scope has to be opened and the optical assembly cleaned or replaced — a procedure from the manufacturer's service documentation, done by authorised personnel. Until that is done, the scope should stay out of clinical use.
Safety
A scope that has been submerged may carry contaminated liquid inside, so keep it quarantined until it passes leak testing, full reprocessing and an electrical safety check before any return to patients.
Lesson: If image artefacts survive lens cleaning after a flooding event, the problem is inside the optics — treat any liquid-damaged endoscope as out of service until it passes leak testing and is properly repaired.
Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.