Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Olympus OER-AW
Olympus · OER-AW

Olympus OER-AW: circulation pump will not start

What happened

The reprocessor aborted its cycle with error E22 and the circulation pump would not start, so no water was being circulated through the chamber.

What was checked

Cause

With the strainer clean and pressure and hoses in order, the engineers found the circulation pump itself had failed and could not start under load.

What was done

The failed pump was replaced. As reported in the case, after the replacement the unit completed its cycle normally and the E22 fault did not return.

Safety

Isolate the reprocessor from the mains before opening the pump compartment, treat all residual and drained water as biohazardous, and confirm a full disinfection cycle passes and the unit passes an electrical safety check before returning it to use.

Lesson: When a circulation fault survives cleaning and supply checks, the flow path is probably fine and the pump itself is the next suspect.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes