Olympus · OER-AW / OER-AV (мойка эндоскопов)

Olympus OER-AW/OER-AV: detergent suction hose breakdown clogging the wash pump

What happened

In an endoscope washer-disinfector, the detergent suction hose delaminated and dissolved, and its debris clogged the wash pump. On machines with several years of intensive use without official service (as reported in the case), the pump and the membrane pump were also worn.

What was checked

Cause

The engineers found that the original suction hose had degraded from prolonged exposure to the detergents and disinfectant solutions, shedding debris into the pump; the pump and membrane pump showed natural wear from years of intensive use without official service.

The cause is probable and was not fully confirmed.

What was done

The engineers replaced the detergent suction hose with a silicone hose compatible with the solutions in use, disassembled and cleaned the clogged pump, and overhauled the membrane pump; a pump motor replacement is still expected on the second machine, as reported in the case. The outcome is reported as probable.

Safety

Before returning the washer to use, isolate the power supply and use chemical protection when opening the fluid lines, then verify with a complete cycle that dosing and disinfection parameters are met, since worn pumps and degraded hoses can leave biohazard-contaminated endoscopes inadequately reprocessed.

Lesson: Hoses and pump wear parts in washer-disinfectors are consumables, so inspect and replace them on a schedule based on material compatibility and hours of use rather than waiting for a failure.

Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.