Olympus X300: xenon lamp at end of life — replacement and counter reset

What happened

The light source needed a xenon lamp replacement and a lamp-hour counter reset.

What was checked

Cause

In this case the xenon lamp had simply reached the end of its rated service life.

The cause is probable and was not fully confirmed.

What was done

The engineers replaced the lamp with the correct type for the unit and then reset the lamp-hour counter using a procedure from the manufacturer's service documentation, done by authorised personnel. They checked that the new lamp was seated correctly and the cooling path was clear before returning the unit to use. The outcome was reported as probable in the case.

Safety

Treat a spent xenon lamp as a sealed glass envelope under high pressure: isolate the unit from mains, let the lamp cool completely, wear eye protection, and transport and dispose of the lamp in its protective sleeve, because a burst lamp can throw glass and damage nearby electronics.

Lesson: Replace a xenon lamp at its rated end of life instead of running it to failure — a lamp that bursts can destroy the light-source electronics and cost far more than the lamp itself.