Opera g500: kV/mA fault during X-ray generation
What happened
On the Opera g500 generator unit of an angiography system, a kV/mA fault appeared during generator operation and interrupted X-ray exposure.
What was checked
- X-ray tube replaced first — the fault stayed (as reported in the case).
- High-voltage cables replaced next — the fault remained (as reported in the case).
- Suspicion then moved to the high-voltage tank; the tank was not the cause, and checks turned to the generator's inverter unit.
Cause
The engineers found a faulty inverter unit inside the generator: its output could not hold stable kV/mA, so the generator raised a kV/mA fault while tube, cables and tank were not to blame.
What was done
The inverter unit was replaced; the kV/mA fault did not return and the generator was reported back to normal operation.
Safety
High-voltage generator circuits and the HV tank can hold a lethal charge after power-off — isolate the unit and verify discharge per the manufacturer's service documentation, and complete electrical safety and X-ray output checks before the system returns to clinical use.
Lesson: When a kV/mA fault survives tube and cable swaps, move step by step through the generator's own power chain — the inverter can be the true source, so confirm each stage by measurement instead of replacing parts on guesswork.
Covered in the course: Angiography and C-arm Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.