Opera T90: jerky column travel

What happened

The column of the X-ray unit moved in strong jerks during travel, and error 031 appeared periodically (reported in the case).

What was checked

Cause

Faulty potentiometer 3pt4: its resistance changed in jumps. Reported in the case, the fault only showed when the potentiometer was measured after removal.

What was done

Replacement of potentiometer 3pt4; the jerking and error 031 did not return after the repair.

Safety

De-energise the system before measuring drive components, and before returning it to use confirm that the column stops at the limit switches and on emergency stop, then run the electrical safety test.

Lesson: A potentiometer can pass a quick in-circuit voltage check and still be faulty, so confirm it by measuring resistance on the removed part.