Pentax · EC38-i10L

Pentax EC38-i10L: image spikes after power cycling

What happened

Sporadic bright spikes/noise on the endoscopic image: the picture is normal at the first power-up, but after the system is switched off and on again, artifacts appear.

What was checked

Cause

The engineers suspected an intermittent break or poor contact in the flat cable of the CCD at the distal end; the cause is probable and was not confirmed by opening the scope.

The cause is probable and was not fully confirmed.

What was done

No repair was completed in the case, and opening or cutting the scope in the field is not an acceptable repair path. Repair or replacement inside the insertion tube/CCD assembly is a procedure from the manufacturer's service documentation, done by authorised personnel; the suspected fix (repair of the CCD cable) remains probable.

Safety

An endoscope with an opened or cut insertion tube loses its seal and can no longer be reliably reprocessed — keep it out of service and return it to use only after authorised repair with a passed leak test, full reprocessing and an electrical safety check.

Lesson: Intermittent image artifacts that appear only after a power cycle usually point to a marginal CCD cable contact — confirm the cause before repair and never cut into a sealed endoscope in the field.

Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.