Pentax · EPK-I5000

Pentax EPK-I5000: LAN interface dead after lightning surge

What happened

After a lightning strike into the building's lightning rod, the network port of the endoscopy system stopped working; this port was used to send endoscopy images over the local network, and other network equipment on the same low-voltage lines was also damaged, as reported in the case.

What was checked

Cause

A voltage surge from the lightning strike travelled into the building over the low-voltage internet line and destroyed the LAN controller chip on the system's motherboard.

What was done

The damaged LAN controller chip was replaced by component-level soldering on the motherboard. As reported in the case, this repair cost around 800 rubles in parts, against a quoted price of 200,000–300,000 rubles for a full motherboard replacement. The network function was restored and the outcome was reported as solved.

Safety

Isolate the unit from mains and network before any board-level work, and after a lightning-related repair complete full electrical safety testing (protective earth continuity and leakage currents) before the system returns to clinical use.

Lesson: A lightning surge reaches medical equipment through low-voltage data lines just as easily as through mains, so network cabling needs surge protection no less than power circuits.

Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →