Philips 3.0T: gradient axes overheat on heavy scans
What happened
On a dual-gradient MRI with a shared chiller/hydraulic cooling loop, the gradient axes went into overheat during heavy-load scans. A more powerful pump was fitted, but flow did not increase — it even got slightly worse — and the hoses at the pump inlet were being squeezed by suction vacuum.
What was checked
- Flow measured on the primary circuit before any pump replacement
- Loop flushed with a citric acid solution
- Full system flushed with stronger cleaning chemistry
- Heat exchanger inspected (the original factory unit)
Cause
The engineers traced the problem to the hydraulic drive in the cooling loop: it was not opening, so the circuit stayed throttled regardless of pump capacity. The collapsing inlet hoses were the clue — the stronger pump was simply pulling harder against this restriction, which is why flow dropped instead of rising.
What was done
The fix was a combination of measures: the loop was flushed clean and the hydraulic drive was repaired. With the drive opening properly, the engineers report the overheating on loaded scans was resolved.
Safety
Before opening any section of a gradient cooling loop, isolate that section and relieve residual pressure — coolant can be hot and the circuit can stay pressurised after shutdown.
Lesson: When a stronger pump does not raise flow and instead collapses the inlet hose, the restriction is downstream — check valves, drives and actuators in the loop before touching the pump again.