Philips · Access

Philips Access: PM lubrication of gantry bearing, collimator and patient table

What happened

Not a breakdown: the engineer asked which lubricants are specified by the maintenance schedule for the gantry bearing, collimator and patient table of a Philips Access CT.

What was checked

Cause

No equipment fault in this case — it was a maintenance-materials question. The risk discussed was old, contaminated or excessive grease: on the collimator and table it can cause stiff movement and jamming, as reported in the case.

What was done

For the gantry bearing, use the grease specified by the manufacturer's maintenance documentation — reported in the case as Mobilgrease 28 (material code 453566361061, as quoted in the case). For the collimator and patient table, completely remove all old grease, apply only a small amount of the fresh specified grease, and do not use abrasive compounds or unspecified lubricants.

Safety

De-energise the scanner and make sure the gantry cannot start or coast before touching the bearing — it stores rotational energy; after greasing, verify collimator and table movement and complete the required functional checks before the system returns to clinical use.

Lesson: Treat grease as a specified material, not a matter of habit: use only the lubricant named in the manufacturer's documentation for each point, renew it fully and apply sparingly — old or contaminated grease is what jams mechanisms.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.