Philips Achieva 1.5T: eddy current test fails with errors jumping between axes
What happened
The Mr eddy current test fails intermittently: one run reports an error on XZ, the next on XY, another on all three axes, with no single consistently failing axis.
What was checked
- Reconnected the interface and coaxial cables inside the gradient cabinet; the failures stayed.
- Reconnected the axis connectors as well; no change.
- Rolled the system back to a backup roughly a year old, as reported in the case; the errors kept moving between axes.
- Took a colleague's suggestion to run a spike test, which pointed to electrical noise rather than one bad axis.
- After that, inspected the magnet module and found the fill hatch not seating tightly.
Cause
A poorly seated fill hatch on the magnet module. The engineers linked the floating errors to the hatch, apparently because the gap let electrical noise into the measurements, which the test showed as errors wandering across the axes.
What was done
Seat and fasten the fill hatch so it closes tight. In this case the eddy current test then passed consistently, and the case is reported as solved.
Safety
Gradient cabinets can hold stored charge in capacitors even after power-down, so follow the manufacturer's isolation and discharge steps before opening them, and confirm every hatch and panel on the magnet enclosure is fully closed before returning the scanner to use.
Lesson: When an eddy current test error keeps jumping between axes, suspect electrical noise from an unsealed cover or hatch before digging into the gradient chain.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.