Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Philips Achieva/Ingenia F2000 (1.0T)
Philips · Achieva/Ingenia F2000 (1.0T)

Philips Achieva/Ingenia 1.0T: F0 frequency shift after gradient coil replacement

What happened

On a 1.0T Philips system with an F2000 teslameter, the reported F0 frequency moved off its nominal value and scans would not run. The engineers found the teslameter reading itself rock-steady, with no drift over time — a one-time offset that had appeared after the gradient coil was replaced.

What was checked

Cause

Reported in the case as most probable: a frequency offset introduced during the gradient coil replacement — not the coil itself, but seating clearances or alignment tolerances during the QBC re-alignment; the root cause was not finally confirmed.

What was done

The gradient coil was replaced with the magnet at field: the coil was rolled into the bore on the standard steel rails supplied for this purpose, over a purpose-made wooden pallet; removing the patient table at field is also possible the same way. Afterwards, the F0 determination and QBC re-centring were carried out as a procedure from the manufacturer's service documentation, done by authorised personnel. In this case the frequency was restored and scanning returned to normal.

Safety

A superconducting magnet at field is a stored-energy and projectile hazard: bring only non-magnetic tools into the room, keep all ferromagnetic items out of the fringe field, and let only personnel trained in MRI magnetic safety work in or near the bore.

Lesson: After a gradient coil replacement, treat F0 as shifted until proven otherwise — re-determine the frequency and re-centre the QBC per the manufacturer's documentation before letting scans run.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes