Philips Achieva: patient table will not feed into the magnet after lowering
What happened
After being lowered, the table would not feed into the magnet any more. Elevation up and down still worked, and the table also stayed dead in manual mode. No error codes were reported.
What was checked
- Full system reboot — behaviour unchanged.
- Fuses under the table checked — all intact.
- Movement tried in manual mode — the table still would not move, so attention shifted from the controls to the wiring.
- The engineers found an open circuit in the table cable right at the flex point: two of the reported 22 conductors were broken.
Cause
Fatigue failure of two of the reported 22 conductors in the table cable, exactly where the cable bends every time the table is lowered. The feed-into-magnet drive lost its signals, while elevation — fed by other conductors — kept working, so one cable produced two very different symptoms.
What was done
The engineers restored the broken conductors and the table fed into the magnet normally again. To stop the fault returning at the same spot, they recommended re-dressing the cable in its drag chain so the flex point moves onto undamaged cable.
Safety
Isolate the table drive before opening the cable, keep tools secured and well away from the fringe field of the magnet, and repeat an electrical safety test (protective earth and leakage current) before the table goes back into use.
Lesson: When one motion works and a neighbouring motion on the same moving cable does not, check the cable at its flex point before blaming drives or boards.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.