Philips Affiniti 70: System hangs at start-up, no boot
What happened
The scanner would not boot: it stopped at an early start-up stage reported in the case as memory initialisation (code 3b) and never reached the operating screen.
What was checked
- Reseated and cleaned the contacts of the memory modules.
- Confirmed the exact motherboard variant before looking for spares.
- Swapped in known-good memory modules from another system.
- Booted with the modules installed one at a time to isolate the faulty one.
Cause
A defective memory module. The engineers found that cleaning contacts and swapping whole modules in pairs did not reveal it; the fault only showed up when each stick was tested on its own.
What was done
Replacing the defective memory module, identified by booting with one module at a time, restored normal start-up.
Safety
Disconnect the system from the mains and follow ESD precautions before handling memory modules, and complete an electrical safety test before returning the scanner to clinical use.
Lesson: When a system hangs during memory initialisation, isolate the fault by testing memory modules one at a time before ordering a motherboard.