Philips Ambition: compressor NTC temperature sensor fault
What happened
On a fairly new scanner, the compressor threw a temperature-sensor fault, and the engineers doubted whether the NTC readings could be trusted at all.
What was checked
- Measure the actual pipe temperature right where the sensor sits and read the sensor's resistance at the same time; the pair of values should match the NTC characteristic for that temperature.
- Establish which sensing technology this compressor variant uses - NTC thermistors or self-resetting bimetallic thermal protectors - because they behave differently and are not interchangeable.
- Before condemning the sensor, check its seating in the pipe well and the connector for poor contact, which can mimic a failed element.
Cause
The engineers found a failed NTC thermistor in the compressor giving implausible readings; the fault was in the sensor itself, not in the circuit behind it. They also noted that some compressor variants carry NTC thermistors while others use bimetallic self-resetting protectors, which are a different kind of part.
What was done
The failed NTC sensor was replaced with a correctly rated part, and the case was closed as solved. Replacing the bimetallic self-resetting protectors with cheaper non-equivalent analogues was rejected: those are protection devices, and any replacement must be like-for-like, done per the manufacturer's service documentation.
Safety
De-energise the compressor and verify zero voltage before touching the sensors - the refrigerant circuit stays pressurised as stored energy, and inside the MR room use only non-ferromagnetic tools with the magnet assumed energised at all times.
Lesson: Prove a suspect temperature reading by comparing the actual pipe temperature with the sensor's resistance before ordering parts, and never downgrade built-in thermal protection to cheaper substitutes.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.