Philips AX700: inspiratory valve sticks again after test/calibration
What happened
After a valve test/calibration the inspiratory valve starts delivering gas into the breathing circuit, but after some time it sticks again; no error codes were reported.
What was checked
- The engineers suspected water or condensate had got inside the inspiratory valve and was making it stick intermittently.
- They proposed removing the valve and flushing it with alcohol while actuating it, so the flush works through the mechanism.
- To make the valve click during flushing, a brief bench supply of about 12 V (reported in the case, per the datasheet) was suggested.
Cause
The most probable cause, reported in the case, was moisture (water or condensate) reaching the inspiratory valve and making it sticky; the exact cause was never confirmed.
The cause is probable and was not fully confirmed.
What was done
The suggested fix, reported as probable, was to flush the valve with alcohol while clicking it with a brief bench supply (about 12 V per the datasheet, as reported), then dry it completely, reassemble and repeat the valve test/calibration; the long-term result was not confirmed.
Safety
Before opening the gas pathway, isolate the machine from mains and from the patient circuit, and do not return it to service until the valve is completely dry of alcohol — flammable vapour in an oxygen-rich circuit — and it passes the manufacturer's leak, self-test and electrical safety checks.
Lesson: A valve that passes a test and then sticks again usually points to contamination such as condensate, so clean the valve mechanism itself rather than just repeating the test.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.