Philips AX700: respiratory rate reading disappears after long runs
What happened
As reported in the case, after about 1.5–2 hours of operation the machine stopped displaying the measured respiratory rate and raised a warning, while continuing to ventilate; the rate value itself did not drift or become erratic before it disappeared.
What was checked
- The engineers found from a screen photo taken during operation that inspired CO2 held around 1.4 instead of the expected 0, and the capnogram curve did not return to the baseline after exhalation.
- Check the CO2 absorbent — in this case it could be exhausted or missing entirely.
- Clarify the ventilation mode: in assisted modes (PSV/VSV) the machine derives respiratory rate from the gas analyzer, while in mandatory VCV it takes it from the flow sensor — this helps separate the two sources.
- checks_ru
Cause
The most probable cause in this case was an exhausted or missing CO2 absorbent: inspired CO2 never returned to zero, and after a while the rate calculation based on the gas analyzer failed in assisted modes, so the display dropped the value with a warning. The diagnosis was probable, not fully confirmed.
The cause is probable and was not fully confirmed.
What was done
The engineers proposed replacing or refilling the CO2 absorbent and then confirming that the inspired CO2 baseline returns to zero. As a temporary workaround, ventilate in a mandatory mode (VCV), where the rate is calculated from the flow sensor and stays displayed.
Safety
A sustained inspired CO2 above zero means the patient is rebreathing: alert the clinical team immediately and make sure the patient is ventilated safely by alternative means before you change or service the absorber.
Lesson: Before blaming the analyzer, read the waveform: an inspired CO2 baseline that never returns to zero points to the absorber, not to the measurement.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.