Philips Brilliance 40: recon fails full scan after system disk swap

What happened

After the reconstructor's system disk was replaced, the recon reported insufficient resources: the scout image worked, but on a full scan the reconstructor refused to reconstruct, complaining about a missing table.

What was checked

Cause

The engineers found that during the disk replacement the required software supplement (addon) had not been installed, so the recon came up incomplete and could not handle full-scan reconstruction.

What was done

The missing software supplement was installed — a procedure from the manufacturer's service documentation, done by authorised personnel. After that the resource error disappeared and full scans reconstructed normally.

Safety

After any work on the recon software, keep the scanner out of clinical use until the manufacturer's post-service checks and quality tests pass, and if you open the recon racks, isolate the power first — server supplies can hold stored charge.

Lesson: When a recon misbehaves straight after a system disk swap, suspect an incomplete rebuild first and walk the manufacturer's full installation sequence, including every software supplement, before blaming hardware or firmware.