Philips Brilliance 64: low readings on 5 V power supply outputs
What happened
On a Philips Brilliance 64 CT, the engineers measured the DC outputs of the rear power supply and got 3.6 / 3.5 / 4.0 V (reported in the case) on three channels labelled 5 V, and suspected the supply was faulty.
What was checked
- Measured the output voltages of the supply modules with a multimeter and compared the readings with the values printed on the units (as reported in the case: 3.6 / 3.5 / 4.0 V on channels labelled 5 V).
- Checked whether such readings are actually abnormal for these DC-DC converter modules by asking experienced colleagues for the typical values for this supply.
Cause
The measured values turned out to be normal for this case: on these power supplies the printed 5 V marking does not match the expected actual reading, and mid-3 V values on these outputs were confirmed as normal. No supply fault was found.
What was done
No repair was needed. The readings were confirmed as normal for these supplies, the unit was left in place and the scanner returned to use.
Safety
Before probing power-supply outputs, isolate the equipment from the mains and allow the internal capacitors to discharge, and take care that meter probes do not short adjacent terminals.
Lesson: A reading below the printed value is not automatically a fault — compare it with the expected actual level from documentation or experienced peers before condemning a power supply.