Philips Brilliance 64: low readings on 5 V power supply outputs

What happened

On a Philips Brilliance 64 CT, the engineers measured the DC outputs of the rear power supply and got 3.6 / 3.5 / 4.0 V (reported in the case) on three channels labelled 5 V, and suspected the supply was faulty.

What was checked

Cause

The measured values turned out to be normal for this case: on these power supplies the printed 5 V marking does not match the expected actual reading, and mid-3 V values on these outputs were confirmed as normal. No supply fault was found.

What was done

No repair was needed. The readings were confirmed as normal for these supplies, the unit was left in place and the scanner returned to use.

Safety

Before probing power-supply outputs, isolate the equipment from the mains and allow the internal capacitors to discharge, and take care that meter probes do not short adjacent terminals.

Lesson: A reading below the printed value is not automatically a fault — compare it with the expected actual level from documentation or experienced peers before condemning a power supply.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.