Philips Brilliance: no remote access to the workstation desktop

What happened

Remote desktop connection to the CT workstation would not go through. The UltraVNC build already installed on the host (1.0.8.2, as reported in the case) refused incoming connections. Installing different remote-access software ended in a blue screen and a software reinstall.

What was checked

Cause

The engineers concluded it was a compatibility/configuration quirk of the installed UltraVNC: the VNC server only comes up when started manually in the user session, so plain incoming connections failed.

What was done

The fix was to keep the proven tool and match versions: download the same UltraVNC version as reported in the case (1.0.8.2), start it manually as a user-level application on the workstation, and only then connect. LiteManager was mentioned as an alternative remote-access tool, but the blue-screen episode shows why new remote software should not go onto the workstation without a good reason and a rollback plan.

Safety

Treat any software install or reinstall on the scanner workstation as a system-level risk: confirm the scanner is not in clinical use, keep a restore path, and verify normal scanner operation before returning the system to service.

Lesson: On a working scanner workstation, match the remote-access setup already proven on that machine (same VNC build, started manually in the user session) instead of installing new software that can take the whole system down.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.