Philips · Brilliance

Philips Brilliance: patient table frame bent after repeated over-travel

What happened

Reported in the case: on a Philips Brilliance CT, the patient table was repeatedly moved by hand closer than its allowed position during cleaning, until the frame was driven hard against the rail and bent. No error codes; the deformation was found on inspection of the frame geometry.

What was checked

Cause

The engineers attributed the damage to the way the room was handled: cleaning staff regularly pushed the table nearer than its allowed position, and in this one case the frame was finally driven into the rail hard enough to bend. This is one case, not a rule.

What was done

The bent frame itself was handled as a repair per the manufacturer's service documentation, done by authorised personnel — a patient table frame is structural and must not be straightened on site by eye. To prevent a repeat, the engineers fitted a simple mechanical stop: an aluminium threshold profile or a low rail fixed to the floor at the closest allowed table position, so the table physically cannot be pushed any nearer. The problem was reported solved.

Safety

A deformed patient table frame is a patient-safety issue: before returning the scanner to use, verify table travel, brakes and geometry against the manufacturer's specifications and complete functional and electrical safety checks.

Lesson: A position limit that depends on people remembering it will eventually be ignored during cleaning — make critical limits physical.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.