Philips · Brilliance

Philips Brilliance: reconstruction RAID keeps dropping disks

What happened

In this reported case, the HDD RAID in the reconstruction subsystem was chronically unstable: disk speeds gradually fell and drives were repeatedly dropped from the array for failing the speed requirements.

What was checked

Cause

The HDDs in the reconstruction RAID had aged to the point where their response times no longer met the array's requirements, so the controller kept dropping them. A contributing role of the power supply was suggested in this case but remained an opinion, not a proven finding.

What was done

As reported in the case, the engineers replaced two to four HDDs in the array with SSDs, and the system has reportedly run without repeat ever since. The practice discussed alongside: OS and applications on SSD, with the page file and bulk data kept on the HDD RAID to spare the SSDs' write endurance. When changing drive types, follow the manufacturer's documentation on supported media for the reconstruction computer.

Safety

Power the reconstruction computer down and isolate it from the mains before swapping drives, and complete the scanner's electrical safety test before it returns to clinical use.

Lesson: Recurring slow-downs and drop-outs in a RAID usually point to worn drives rather than a controller fault — replace the media in time and match it to the write load.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.