Philips Brilliance (SW 4.1.7): remote-access installers blocked on workstation
What happened
On the workstation of a Philips Brilliance CT (the case reports software version 4.1.7), third-party remote-access tools such as TeamViewer and AnyDesk would not install — the engineers found the setup files being stopped by the McAfee Solidcore agent on the workstation.
What was checked
- Confirm what actually stops the installer: in this case the engineers found the McAfee Solidcore application-control agent blocking third-party setup files before they ran.
- Test whether safe mode changes anything — reported in the case that it did not, which points to an agent-level block rather than a Windows problem.
- Check the manufacturer's service documentation for the approved way to enable remote support on this software release before changing anything.
Cause
By design, the Solidcore application-whitelisting agent allows only approved software on the scanner's workstation, so third-party remote-access installers are treated as unauthorised changes and blocked.
The cause is probable and was not fully confirmed.
What was done
The outcome in this case was reported as probable: the fix was a procedure from the manufacturer's service documentation, done by authorised personnel. Disabling or otherwise evading the protection agent, or using service-level accounts to slip software past it, is not an acceptable route on a medical device.
Safety
After any software change on the workstation, confirm the scanner powers up and scans normally and pass an electrical safety check before return to clinical use, and keep any remote-access channel closed when it is not needed.
Lesson: A protection agent blocking unapproved software on a medical workstation is the safeguard working as designed, so the way forward is the manufacturer's documented procedure, not a workaround.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.