Philips · Chorus 1.5T

Philips Chorus 1.5T: random breaker trips on the gradient amplifier cabinet

What happened

The supply breaker on the gradient amplifier cabinet tripped at irregular moments, mostly towards the end of a scan; after resetting, scanning continued normally. The amplifier's own fault logs were empty and its built-in diagnostics were not available.

What was checked

Cause

In this one case the engineers found two combined faults: an intermittent contact on the Z-mode switch of the upper signal-former board (a cold solder joint was suspected) that moved with the switch actuator, and a contaminated cooling loop for the amplifier with poor flow and an air lock in the expansion tank. The single error 134 was reported in the case as a +15 V dip in the small gradient amplifier; checking its electrolytic capacitors was advised.

What was done

The contact fault on the Z-mode switch was corrected by a procedure from the manufacturer's service documentation, done by authorised personnel. The amplifier cooling loop was flushed with a citric acid solution, the air lock in the expansion tank was cleared, the coolant was topped up, and a digital flow indicator was fitted — it read 25 L/min, reported in the case against an expected 40 L/min. A change of the coolant from distilled water to a 50/50 ethylene glycol mixture was planned.

Safety

Before opening the gradient amplifier cabinet, isolate the supply and allow internal capacitors to discharge, and after any cooling work verify the loop is leak-free with flow within specification before returning the scanner to use.

Lesson: When a breaker trips with no pattern and no logs, suspect a floating contact combined with marginal cooling: correlate the trips with operating conditions and measure the cooling flow before blaming the electronics.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.