Academy / Field fault notes / Philips ClearVue 550
Philips · ClearVue 550

Philips ClearVue 550: fetal size and weight calculations not produced

What happened

Clinicians reported that during obstetric scans the unit would not output the calculations comparing fetal size and weight to gestational age, indicating that the needed settings were not available.

What was checked

Cause

Not confirmed in this case. The engineers suspected that patient parameters (LMP, dates) were not entered at patient creation, or that the system date had drifted, or that the obstetric calculation options were not present on the unit; diagnostics were still in progress when the case was shared.

The cause is probable and was not fully confirmed.

What was done

The recommended steps were to enter full patient data (LMP, gestational dates) when creating each new patient and to check the options list via Setup → Options. The outcome was reported as probable: if the required option turns out to be missing, enabling it is a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

After any settings or configuration work, complete the electrical safety checks from the manufacturer's documentation before returning the scanner to clinical use, and keep standard probe disinfection between patients.

Lesson: Before suspecting a fault, rule out operator-level inputs and the installed options — a missing calculation is often just unfilled patient data, and option availability should be verified before ordering parts.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.