Philips CT: image artifacts on large-FOV scans

What happened

Artifacts on CT images with no error codes. Per the technologists, artifacts often appeared when the unit had not been warmed up. They were visible at large FOV, and at small FOV when the scan plane was offset above or below the gantry axis. The engineers also noted a repeating pattern with a spacing reported at about 36–39 mm, though its origin was not confirmed in the case.

What was checked

Cause

From the survey and full-width scans, the engineers suspected a fault in a detector channel at the edge of the detector row. The pattern — artifacts at large FOV and at offset small-FOV positions — fits an edge-detector problem. This remained a probable cause, not confirmed by a part replacement.

The cause is probable and was not fully confirmed.

What was done

No repair was completed in the case. The conclusion was that when several calibrations fail to clear the artifacts, further detector diagnostics are required per the manufacturer's service documentation.

Safety

During test scans keep radiation exposure controlled: room clear of unauthorised staff, dose per protocol, and image QA completed before the scanner returns to clinical use.

Lesson: When artifacts survive several consecutive calibrations — especially at large FOV — stop re-calibrating and move to detector diagnostics.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.