Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Philips CX50
Philips · CX50

Philips CX50: urology presets missing after software update

What happened

On a portable CX50 after a software update (reported as version 5.0.2) with the urology option enabled, the expected urology presets did not appear in the preset list. No error codes were reported.

What was checked

Cause

The engineers found that the C10-3v is a gynecological endocavitary probe and has never carried prostate presets. Presets on this system are tied to the specific transducer, so with this probe connected the urology list stays empty regardless of the enabled option — by design, not a fault.

What was done

No repair was needed. The recommendation was to create a custom user preset for prostate scanning on the C10-3v manually — reported as quick to do — or, if standard urology presets are required, to use a transducer that supports them according to the manufacturer's documentation. Case closed as solved.

Safety

After any software update, or when working with custom presets, verify before returning the unit to clinical use that it starts normally, the probe is recognised, and the MI/TI acoustic output indices are displayed and behave sensibly on the preset you created.

Lesson: Presets follow the transducer, not the option — before chasing a software fault, check that the probe in use is actually meant to carry the presets you are looking for.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes