Philips Digital Diagnost: collimator light source swapped from halogen to LED
What happened
A routine replacement of the collimator halogen lamp, done in this case as a swap to an LED; voltage, current and brightness had to be matched to the existing optics.
What was checked
- The engineers measured the LED voltage and current; the drive current was set, as reported in the case, to about 0.5 A.
- Diode temperature was checked after a run of some 15 minutes, reported at around 52 °C.
- The diode was selected so that its emitting die area roughly matched the halogen filament area, keeping compatibility with the existing optics.
- Feedback on the color temperature of the new light was collected from the clinicians working with the unit.
Cause
No fault as such: a planned lamp replacement. The engineering task was finding an LED that the existing optics accept and keeping its heat under control.
What was done
In this case the engineers installed a Cree XHP70 LED (12 V, 4500 K) in place of the halogen lamp and set the current to about 0.5 A, reported as roughly a quarter of the diode's capability — enough light with modest heat. Their cooling notes from the case: at 20–25 % of diode power a small heatsink is enough, at around 50 % add a heatsink and preferably a fan, and at full power forced cooling is a must. Outcome reported as solved, with the clinicians satisfied by the white light.
Safety
The collimator light field stands in for the X-ray field, so after any change of the light source verify the alignment of the light and radiation fields per the manufacturer's quality-control procedure, and run the electrical safety checks before returning the unit to clinical use.
Lesson: When replacing a collimator lamp with an LED, success depends less on the diode itself than on matching its emitting area to the original optics and sizing the cooling with generous margin.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.