Philips Eleva: DICOM Worklist stopped arriving while connection test passes
What happened
A Philips Eleva digital radiography system stopped receiving the DICOM Worklist from the facility's RIS, although the connectivity and availability test to the DICOM node reported success.
What was checked
- Ran the DICOM node connection and availability test — it passed, so the network path looked healthy.
- Checked the facility network for a duplicate IP address on the modality — no conflict was found in this case.
- Reviewed the DICOM Worklist settings on the system — no signs of a local misconfiguration.
Cause
The engineers found that the fault was on the RIS side: the server accepted connections, but its Worklist service was not delivering data, so the passing connectivity test gave a false picture of a healthy chain.
What was done
The RIS-side problem was resolved by the responsible IT staff; once it was fixed, the worklist started arriving again and the system returned to normal use. No work was needed on the modality itself beyond verification.
Safety
Before disconnecting or re-terminating any network cabling on the modality, power down and isolate the equipment according to local practice, and after any network change confirm with a non-clinical check that the system behaves normally before returning the room to patient use.
Lesson: A passing DICOM connectivity test only proves the network path, not the service — when the test succeeds but the worklist stays empty, expand the search beyond the modality to the RIS side.