Philips Epiq 5: workstation array corrupted during disk imaging attempt
What happened
While taking a disk image of the workstation drive with a disk-imaging utility, the on-board semi-hardware Intel chipset RAID controller was detected as two separate disks instead of one array; the tool read and wrote across both, and data on the array was damaged.
What was checked
- The engineers checked how the Intel chipset RAID controller was presented to the imaging utility.
- Before any disk work on such a workstation, verify that the software sees the array as a single volume, not as individual drives.
Cause
The imaging utility version in the case had no drivers for the newer Intel chipset RAID controller, so it did not recognise the array and treated the member disks as two independent drives; its operations destroyed the array data. This was found in one case, not a general rule.
What was done
The system had to be fully reinstalled from scratch. As advice reported in the case: before imaging a RAID-backed workstation, use an up-to-date server edition of the imaging software with drivers for the current chipset RAID controllers, or a sector-level disk editor that handles the array correctly.
Safety
Power the workstation down and disconnect it from the mains before opening it to access the drives, take ESD precautions, and after a full system rebuild complete the manufacturer's verification checks before returning the scanner to clinical use.
Lesson: On a RAID-backed system, always confirm that the backup or imaging tool actually sees the array as one volume before letting it touch the disks — otherwise the backup attempt itself can destroy the data it was meant to protect.