Philips Epiq (PureWave probe): lens delamination with color Doppler noise
What happened
On a Philips Epiq with a PureWave probe, the factory lens had delaminated from the probe face. In color Doppler mode this showed as strong noise across the whole scanning field, reported in the case as most noticeable at gain above 40.
What was checked
- Inspect the probe face and lens for delamination: gaps, bubbles or lifted edges that break the acoustic path and acoustic coupling.
- Run the probe in color Doppler on a known target or phantom to confirm the noise comes from the probe itself, not from the main unit or another probe port.
Cause
The exact mechanism was not established in this case, but the engineers found that a delaminated lens is a likely source of the color Doppler noise: an air gap or damaged lens layer disrupts the acoustic path and degrades the signal.
What was done
The fix was a lens replacement (re-bonding) on the probe, done as a procedure from the manufacturer's service documentation by authorised personnel. After the replacement the probe returned to full service and the noise was gone, reported in the case.
Safety
Before returning a probe that has had lens work to clinical use, run an electrical safety test (leakage current check) per the manufacturer's requirements and inspect the lens surface for residual damage or trapped disinfectant.
Lesson: Whole-field color Doppler noise can come from a physical probe defect such as lens delamination, so always inspect the probe and its acoustic path before troubleshooting the system.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.