Philips HD11: system will not start from the panel power button
What happened
The scanner would not start when the power button on the front panel was pressed, even though parts of the embedded computer appeared to run when powered manually.
What was checked
- The engineers first checked manual start of the computer power supply; outside the system it started on its own.
- They checked for the relay click when the panel power button was pressed.
- With the supply started manually, they confirmed the hard drive and fans could operate.
Cause
The engineers found the fault in the embedded computer's power supply. This is a known weak point on this model: the supply can bench-start fine outside the system, but fails to work as part of the running computer.
The cause is probable and was not fully confirmed.
What was done
The computer power supply was removed and taken for replacement or repair. The outcome was reported as probable: a working supply was expected to restore normal start-up from the panel button, but confirmation would only come when the unit was powered back up.
Safety
Disconnect the system from mains and allow internal capacitors to discharge before opening the power supply enclosure, and complete a full electrical safety test before returning the scanner to clinical use.
Lesson: A power supply that starts fine on the bench can still be the fault — always test how it behaves inside the system under the computer's start signal before blaming anything else.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.