Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Philips HD15
Philips · HD15

Philips HD15: no power-up, then image not filling the screen

What happened

The scanner would not power on at all. After the settings memory was cleared and the backup battery replaced, it started and worked, but the live image occupied only part of the screen.

What was checked

Cause

The no-start was most probably a flat BIOS backup battery with lost CMOS contents — the unit came back after the battery was replaced and the memory cleared. The small picture was not a fault at all: the engineers found that this is normal behaviour of the model, with the picture size set by a documented option.

What was done

The engineers replaced the BIOS backup battery and cleared the settings memory, and the scanner powered up. For the screen size, they found the picture setting described in the user manual and set it as needed; no repair was required for that part.

Safety

Before opening the cabinet to reach the board battery, isolate the scanner from the mains and take anti-static precautions; carry out the electrical safety test before returning the unit to clinical use.

Lesson: After a settings reset or an odd-looking picture, read the manual first — what looks like a fault can be normal, adjustable behaviour.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes