Philips HD15: periodic image doubling with the endocavitary probe
What happened
On a Philips HD15 the image periodically doubles when scanning with an endocavitary probe; the fault is intermittent and could not be reproduced during the initial inspection.
What was checked
- The engineers bent the cable at the probe tail/handle to try to reproduce the fault; this gave nothing.
- The plan reported in the case: flex and twist the cable along its entire length, including a twisting mode, to catch the intermittent break.
Cause
Suspected, not confirmed: broken or damaged conductors inside the probe cable, making intermittent contact as the cable moves.
The cause is probable and was not fully confirmed.
What was done
Reported in the case as the planned fix: flex and twist the cable by hand section by section along its full length to localize the intermittent break; if damage is confirmed, the probe is repaired or replaced per the manufacturer's service documentation, done by authorised personnel. The outcome was probable, not yet proven.
Safety
An endocavitary probe is an invasive patient-applied part: with suspected internal cable damage, take it out of service and confirm insulation and leakage-current checks pass before returning it to use, keeping the approved reprocessing routine.
Lesson: An intermittent image fault with one probe points first to its cable — flex and twist the whole cable section by section, not only the tail at the handle.