Philips HD15: stopped starting up, restarts no longer help
What happened
Reported in the case: the unit stopped starting up. Earlier it would come up after several restarts — five or six, as reported — but now it does not start at all. No error codes were reported.
What was checked
- Check whether the HV voltage from the power supply is actually present.
- Inspect the ribbon cable from the PSU to the vertical board: verify seating and contact, and re-form (bend) the contacts if they look loose or flattened.
- Pull the logs from the hard drive for boot and shutdown clues.
Cause
Not confirmed in this case. A probable cause is a poor, intermittent contact in the ribbon cable from the power supply combined with a missing HV rail, by analogy with a similar case on an HD11 where exactly this combination was found.
The cause is probable and was not fully confirmed.
What was done
Recommended in the case: verify HV presence and re-form/re-seat the PSU ribbon cable contacts. The fix was not yet confirmed at the time of the case — checking HV, contacts and logs was still ahead.
Safety
Unplug the system from mains and let the power supply discharge before touching its connectors or probing for HV — the HV section can hold a dangerous charge — and run an electrical safety test before returning the unit to clinical use.
Lesson: When a fault that used to clear with a few restarts finally stops responding, look first for an intermittent connection or a drooping supply rail — such degradation only gets worse.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.