Philips · HD15

Philips HD15: system does not start from the power button

What happened

The ultrasound system would not start from the front power button; when the internal system unit (PC) was powered separately, it started normally.

What was checked

Cause

Probable, not confirmed in the case: the standby section of the main power supply had failed, so pressing the start button woke nothing. The engineers also considered that the PSU was not feeding the upper connector of the inverter, or that the electrolytic capacitors in the computer's ATX unit had dried out.

The cause is probable and was not fully confirmed.

What was done

No confirmed repair is recorded in the case. The working conclusion pointed to the main power supply — its standby section, or dried-out capacitors in the computer ATX unit. The probable fix was repair or replacement of that PSU, a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Switch-mode power supplies hold a lethal charge in their capacitors after the mains is removed: isolate the unit, wait, and verify the DC bus is discharged before touching anything, and run a full electrical safety test before returning the system to clinical use.

Lesson: When a system boots with its PC fed separately but not from the power button, check the main PSU's standby and control outputs first — the start chain lives there, not in the computer itself.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.