Philips Ingenuity / Family One: question on slip ring and brush lubrication at service
What happened
During planned maintenance the engineers questioned whether the slip ring and brushes on this CT should be lubricated at all, and what the correct procedure is for this model.
What was checked
- Checked the manufacturer's service documentation for the model-specific slip ring maintenance schedule.
- Identified the type of brushes installed on the unit — graphite or metal, since the lubrication requirement depends on this.
Cause
In this case the engineers found that on the Ingenuity ring lubrication is part of the maintenance schedule, while on the Family One with graphite brushes greasing is explicitly wrong: grease makes the graphite dust stick to rings and brushes instead of falling away, which worsens contact over time.
What was done
On the Ingenuity the rings were lubricated with a precisely measured dose applied with a microliter syringe, the amount matched to the ring size as prescribed in the manufacturer's documentation; on the Family One no lubrication was applied and the dry brushes were left as they are. Reported in the case as solved.
Safety
Before any work at the slip ring, fully power down the gantry and follow the manufacturer's lockout and discharge procedure: slip ring circuits can carry high voltage and rotating parts can move unexpectedly.
Lesson: Never assume lubrication helps a slip ring — the correct action depends on the brush material and the model, so always follow the model-specific service documentation before touching rings or brushes.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.