Philips · Intera 1.0T

Philips Intera 1.0T: degraded spine coil images and knee coil failure after cryo service

What happened

After a displacer and adsorber replacement, image quality on the spine coil degraded and the knee coil would not scan. No error codes were reported in the case.

What was checked

Cause

The timing misled everyone: earlier that day a lamp in the room had been replaced during maintenance, and the new lighting introduced interference into the images. The cryogenic work itself was not the cause — a coincidence of two unrelated events in one day.

What was done

The interference source was removed. Turning the lights off served as the test; eliminating the noise from the replaced lighting restored image quality on the spine coil and the knee coil scanned normally again. No parts were changed in this case.

Safety

Only MR-compatible lighting and fixtures belong in the scan room, and anything brought inside — even for a quick lamp change — must pass the facility's ferromagnetic screening before entering the static field.

Lesson: When image artefacts appear right after unrelated service work, test the simplest interference sources first — switching lights and nearby devices off takes minutes and can save hours of cryogenic and subsystem diagnostics.

Covered in the course: MRI Service — Level 2: Fault-Finding →