Philips Intera 1.0T: degraded spine coil images and knee coil failure after cryo service
What happened
After a displacer and adsorber replacement, image quality on the spine coil degraded and the knee coil would not scan. No error codes were reported in the case.
What was checked
- The engineers asked what, if anything, the system reported when the knee coil failed; nothing conclusive came up in the case.
- Magnet (cryostat) temperature was considered a suspect by analogy with similar cases on other platforms.
- The decisive check was simply switching the room lights off as an interference test — the image artefacts cleared, pointing away from the cryogenic work.
Cause
The timing misled everyone: earlier that day a lamp in the room had been replaced during maintenance, and the new lighting introduced interference into the images. The cryogenic work itself was not the cause — a coincidence of two unrelated events in one day.
What was done
The interference source was removed. Turning the lights off served as the test; eliminating the noise from the replaced lighting restored image quality on the spine coil and the knee coil scanned normally again. No parts were changed in this case.
Safety
Only MR-compatible lighting and fixtures belong in the scan room, and anything brought inside — even for a quick lamp change — must pass the facility's ferromagnetic screening before entering the static field.
Lesson: When image artefacts appear right after unrelated service work, test the simplest interference sources first — switching lights and nearby devices off takes minutes and can save hours of cryogenic and subsystem diagnostics.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.