Philips iU22: no start — AC relay drops out seconds after power-on
What happened
On power-up the AC-line relay clicks in, then drops back out a few seconds later (reported in the case as about four seconds): the power supply apparently never delivers the feedback signal needed to start the system.
What was checked
- Measure the 48 V rail at the power supply output, as suggested in the case
- Check the remaining power supply output voltages against the values in the service documentation
Cause
The suspected cause is a missing 48 V rail: without it the start-up feedback never appears, so the system drops the AC relay again. In this single case the cause was not confirmed.
The cause is probable and was not fully confirmed.
What was done
The recommended step in this case was to confirm the 48 V output and, if the rails are out of range, have the power supply output voltages adjusted — a procedure from the manufacturer's service documentation, done by authorised personnel. As reported, the result had not yet been confirmed, so treat this as the probable fix, not a proven one.
Safety
Isolate the system from the mains and let the power supply capacitors discharge before probing inside, and carry out an electrical safety test before returning the unit to clinical use.
Lesson: When the AC relay drops out seconds after power-on, check the power supply rails and their start-up feedback signal before suspecting anything else.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.