Philips Juno DRF: software starts then closes with an error

What happened

In this case the control software of the PID computer started, then reported an error and closed by itself. At first it looked like a failing system disk, but the disk was not the cause.

What was checked

Cause

Poor contact of the boards inside the PID computer: at startup the software hit the unstable connection and shut down with an error.

What was done

The engineers eliminated the poor board contact in the PID computer, after which the software started and ran normally. The outcome in the case was reported as solved.

Safety

De-energise and isolate the system before opening the PID computer, and return an X-ray unit to clinical use only after the required electrical safety and functional checks are completed.

Lesson: Startup crashes that mimic a dying disk can come from something as plain as poorly seated boards, so check mechanical contacts before swapping parts.