Philips MicroDose L30: system won't start, only the chiller fan runs

What happened

After assembly the mammography system refused to start up. The only sign of life was the thermochiller fan spinning; no error codes were reported in the case.

What was checked

Cause

Both rack power supplies were dead, so the control electronics had no power at all, while the chiller ran on its own feed and kept spinning its fan.

The cause is probable and was not fully confirmed.

What was done

The fix reported in the case was to test and replace both failed rack power supplies. The outcome was treated as probable, since the diagnosis relied on symptoms matching a previously seen case, not on direct confirmation.

Safety

Isolate the cabinet from the mains and allow the power supply capacitors to discharge before working inside the rack, and complete the electrical safety checks (protective earth and leakage currents) before returning the system to clinical use.

Lesson: When an assembled system shows almost no signs of life, verify every power supply rail at its source before chasing individual subsystems.