Philips MobileDiagnost wDR: unit does not see the battery and blocks exposure
What happened
A mobile X-ray unit did not detect its battery pack and would not allow an exposure to be taken.
What was checked
- The fault was reviewed in a remote consultation with a colleague
- History confirmed: the battery had just been replaced before the problem appeared
Cause
The engineers found that the freshly fitted battery pack was not recognised by the unit until it had been charged: in this case the pack needed a period on mains charging after replacement before the unit would accept it.
What was done
As reported in the case, the engineers left the unit connected to mains power for the roughly 12 hours stated in the case; after that the unit detected the battery and exposures worked again. Outcome: solved.
Safety
Isolate the unit from mains before battery replacement, and before returning it to clinical use verify correct battery indication and perform a test exposure under local radiation safety procedures.
Lesson: After a battery replacement, an unrecognised pack may simply need its initial charge on mains — give the new battery time to charge before hunting for deeper faults.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.