Philips / Siemens MRI: FOV and matrix not adjustable as expected

What happened

In one reported case, engineers comparing two consoles noticed that scan geometry behaved differently: on the Philips system the FOV could be typed in manually over a wide range, while on the Siemens system the shape of the scan region seemed locked and the geometry tab looked different from what they expected.

What was checked

Cause

No fault was found on either system. The vendors simply implement geometry differently: on Philips the FOV can be entered manually in a wide range, while on Siemens the shape of the scan region is governed by the software version — older versions show the geometry tab with a different layout, which makes the option look missing.

What was done

No repair was needed. The engineers worked within each console's own logic: set the FOV manually where the system allows it, changed the first two matrix values (frequency and phase-encoding direction) and left the third value untouched, and on the Siemens system accepted that the scan-region shape follows the software version. Any software update itself is a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Any change to scan geometry is still a protocol change: keep it within the limits approved for the site, and let only authorised personnel touch system software — never try to unlock version-dependent options yourself.

Lesson: Scan geometry options are vendor- and software-version specific, so check what the console actually offers before calling it a fault.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.