Philips / Siemens MRI: FOV and matrix not adjustable as expected
What happened
In one reported case, engineers comparing two consoles noticed that scan geometry behaved differently: on the Philips system the FOV could be typed in manually over a wide range, while on the Siemens system the shape of the scan region seemed locked and the geometry tab looked different from what they expected.
What was checked
- Check on each console whether the FOV can be entered manually and within what range, as reported in the case.
- Open the geometry tab and note its layout and available fields; its appearance changes with the software version.
- When adjusting the matrix, plan to change the first two values (frequency and phase-encoding direction) and leave the third value alone.
Cause
No fault was found on either system. The vendors simply implement geometry differently: on Philips the FOV can be entered manually in a wide range, while on Siemens the shape of the scan region is governed by the software version — older versions show the geometry tab with a different layout, which makes the option look missing.
What was done
No repair was needed. The engineers worked within each console's own logic: set the FOV manually where the system allows it, changed the first two matrix values (frequency and phase-encoding direction) and left the third value untouched, and on the Siemens system accepted that the scan-region shape follows the software version. Any software update itself is a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Any change to scan geometry is still a protocol change: keep it within the limits approved for the site, and let only authorised personnel touch system software — never try to unlock version-dependent options yourself.
Lesson: Scan geometry options are vendor- and software-version specific, so check what the console actually offers before calling it a fault.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.