Planmed Melody 2: error message on exposure attempt
What happened
On a Planmed Melody 2 mammography unit, an error message appeared whenever an exposure was attempted, and no image could be taken.
What was checked
- Photographed the display before pressing the exposure button to record the selected operating mode.
- Compared the selected AEC curve with the curves actually programmed in the unit: the selected one was not programmed.
Cause
The engineers found that an AEC curve that had never been programmed had been selected on the console; without valid curve data the unit blocked the exposure with an error message.
What was done
No parts and no service access were needed: the operators were shown how to switch back to the standard, factory-programmed AEC mode through the normal user interface, and the unit exposed normally again.
Safety
Because the fault affected exposure control, confirm correct behaviour with a test exposure and check the mAs/dose indication before returning the unit to clinical use, following local radiation safety procedures.
Lesson: An exposure error is not always a hardware fault: before opening anything, check which mode and AEC curve the console has selected, because an unprogrammed user selection can block exposures exactly like a real failure.
Covered in the course: Mammography Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.