Plasma 100: hydrogen peroxide cannot be drained from the menu
What happened
The sterilizer needed its hydrogen peroxide sterilant removed from the system, and the engineers looked for a drain item in the menu but could not find one.
What was checked
- Searched the operator menu for a sterilant drain function — as reported in the case, no such item exists on this model.
- Traced the sterilant path (bottle line, drain pump, drain point) to plan a controlled manual drain into a compatible container.
Cause
In this case the model simply has no menu routine for emptying the sterilant: peroxide removal is a physical service operation, and any drain driven by the unit's own expiry logic is a manufacturer-level function.
What was done
The engineers drained the peroxide mechanically: through the drain line with the drain pump disconnected into a suitable container, or by drawing the sterilant out through the bottle tube. No settings or protections were touched; if a drain triggered by the unit's expiry logic is needed, that is a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Hydrogen peroxide is a corrosive oxidiser: isolate the sterilizer from the mains before opening it, wear eye/face protection and chemical-resistant gloves, drain into a compatible vented container in a ventilated area, and verify leak-free reassembly and correct refilling before return to use.
Lesson: No menu function does not mean no job: know the fluid path, drain manually with proper peroxide handling, and never use workarounds that exploit the unit's protective logic — stick to the manufacturer's documented procedures.
Covered in the course: Steriliser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.