Plaster 120: sterilizer door won't close at power-up
What happened
Plasma sterilizer door stayed open when the unit was switched on and would only close after a couple of power cycles.
What was checked
- Checked the door limit switches and their actuation
- Measured the voltages on the door control board near the trimmer potentiometers
Cause
The engineers found low voltage on the door control board near the trimmers — reported in the case as 0.19 V instead of the expected value. With the reference that low, the board did not drive the door reliably at power-up, and only a restart happened to bring it to life.
What was done
The engineers adjusted the trimmer potentiometer on the door control board, bringing the voltage to 0.25 V as reported in the case. After that the door closed normally at power-up, with no restarts needed. No parts were replaced.
Safety
Isolate the sterilizer from the mains and let internal capacitors discharge before touching the control board. The door is a safety interlock: after any adjustment, confirm it locks and its switches operate correctly, and run an electrical safety test before returning the unit to service.
Lesson: When a door only works after restarts, check the reference voltages on its control board before blaming switches or mechanics.
Covered in the course: Steriliser Service — Level 2: Fault-Finding →