ПроСкан · Проскан-2000

ProScan-2000: vertical stripes across the whole image after transport

What happened

After the unit was transported, digital images showed vertical stripes over the entire field, with no error codes reported in the case.

What was checked

Cause

No root cause was proven. The most likely explanation, given that a full realignment cured the defect, is that shocks during transport knocked the scanning detector alignment out.

What was done

With no mechanical fault confirmed, the engineers went back to the beginning and carried out the complete alignment of the scanning linear detector, step by step, following the manufacturer's service documentation. Almost all the stripes disappeared and the case was closed for the time being, with the unit left under observation.

Safety

High-voltage circuits of the generator can hold a lethal charge after switch-off; follow the discharge and verification steps in the service manual, and confirm image quality and dose with test exposures before the unit returns to clinical use.

Lesson: After transport, suspect alignment and calibration of a scanning detector first and run the full manufacturer's alignment procedure from the start, instead of stopping at the first plausible fault.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.