ProScan-2000: vertical stripes across the whole image after transport
What happened
After the unit was transported, digital images showed vertical stripes over the entire field, with no error codes reported in the case.
What was checked
- Cleaned the guide rails and checked that the scanning carriage ran freely; the stripes stayed as before.
- Re-tightened the plastic gear on the stepper motor shaft; no improvement was reported in the case.
- Checked the detector histogram; the peak sat in the centre, so the readout chain looked plausible.
- The engineers suspected a high-voltage breakdown and discussed running the generator into a dummy load to separate generator from tube faults; this turned out to be unnecessary.
Cause
No root cause was proven. The most likely explanation, given that a full realignment cured the defect, is that shocks during transport knocked the scanning detector alignment out.
What was done
With no mechanical fault confirmed, the engineers went back to the beginning and carried out the complete alignment of the scanning linear detector, step by step, following the manufacturer's service documentation. Almost all the stripes disappeared and the case was closed for the time being, with the unit left under observation.
Safety
High-voltage circuits of the generator can hold a lethal charge after switch-off; follow the discharge and verification steps in the service manual, and confirm image quality and dose with test exposures before the unit returns to clinical use.
Lesson: After transport, suspect alignment and calibration of a scanning detector first and run the full manufacturer's alignment procedure from the start, instead of stopping at the first plausible fault.
Covered in the course: X-ray Service — Level 2: Fault-Finding →